Dental X-ray Imaging System GDP-1C
ManualEFM6TVBSJP
This Product is Medical Equipment.
The specification above can be changed to improve performance.
]
GDP-1 & GDP-1C
Copyright© 2012 2013 / 5 / 31 Document Ver. 0.1
Table of Contents
TablCFG%QPtenKFQs Introducion CN Manual ...................................................................................................................................................................................... 4 CertificaFQ FG)CTCPVsC ......................................................................................................................................................................................5 SsmbolQU usCFQUGPGN productQ.................................................................................................................................................................. 6 #FXGTVGPEKCUCPVGUFGNWUQ.............................................................................................................................................................................. 7 5GIWTKFCF'NoEVTKEC............................................................................................................................................................................................... 8 5GIWTKFCFFG4CFKCEKxP.................................................................................................................................................................................... 10 1.
GDP-1C Introducion....................................................................................................................................................... 11 1.1
Intended Use .................................................................................................................................................................................. 11
1.2
Caracteristicas ................................................................................................................................................................................ 11
1.3
System Basic Specification ...................................................................................................................................................... 12
1.4
Equipment Composition ........................................................................................................................................................... 13
1.5
Accessories ...................................................................................................................................................................................... 14
1.6
Patient Apply Part........................................................................................................................................................................ 15
1.7
On / Off Switch ............................................................................................................................................................................. 16
1.8
Hand Switch .................................................................................................................................................................................... 17
1.9
Emergency Stop Switch............................................................................................................................................................. 18
1.10 Equipment’s Control Button ................................................................................................................................................... 19 1.11 Laser Control Knob...................................................................................................................................................................... 20 1.12 Changing Detector ...................................................................................................................................................................... 21 1.13 Device status LED ......................................................................................................................................................................... 22 2.
2
OP (Operation panel) .............................................................................................................................................................................. 23 2.1
OP Structure.................................................................................................................................................................................... 23
2.2
X-ray Exposure Mode ................................................................................................................................................................ 25 2.2.1
Panorama program......................................................................................................................................................... 25
2.2.2
Cephalo Program ............................................................................................................................................................. 29
2.3
Choose Patient Size..................................................................................................................................................................... 31
2.4
Exposure Speed and Change of Image Quality ............................................................................................................ 31
2.5
Selection of Jaw shape .............................................................................................................................................................. 31
2.6
X-ray exposure condition change........................................................................................................................................ 32
2.7
Order list ........................................................................................................................................................................................... 32
2.8
Device condition display screen ........................................................................................................................................... 34
2.9
Option setting ................................................................................................................................................................................ 35 2.9.1
General .................................................................................................................................................................................. 36
2.9.2
DB Server ............................................................................................................................................................................. 37
2.9.3
DB Server ............................................................................................................................................................................. 38
2.9.4
Version .................................................................................................................................................................................. 39
PAPAYA PLUS User manual
Table of contents 3.
Image acquisition ...................................................................................................................................................................................... 40 3.1
Registration of patients and orders. .................................................................................................................................. 40
3.1.1 Registration with Triana software. ...................................................................................................................................... 40 3.1.2 Registration with OP software. ............................................................................................................................................. 41 3.2
Panorama mode Set position of patient.......................................................................................................................... 42
3.3
TMJ mode patient position set............................................................................................................................................. 46
3.4
Sinus mode patient position set .......................................................................................................................................... 51
3.5
Cephalo mode patient position set .................................................................................................................................... 56
3.6
Image taking................................................................................................................................................................................... 58
3.7
Image Control ................................................................................................................................................................................ 60
3.8
TWAIN INTERFACE....................................................................................................................................................................... 61
3.9
OP Additional Function.............................................................................................................................................................. 61 3.9.1
Sending to DICOM Server........................................................................................................................................... 62
3.9.2
Burning CD ......................................................................................................................................................................... 62
3.9.3
Register addition order ................................................................................................................................................. 63
3.9.4
Opening of patient's folder ......................................................................................................................................... 63
4.
Cleaning .......................................................................................................................................................................................................... 65
5.
Handling of waste ..................................................................................................................................................................................... 65
6.
Message.......................................................................................................................................................................................................... 66
7.
6.1
Error message ................................................................................................................................................................................ 66
6.2
Condition message ...................................................................................................................................................................... 68
Technical Data ............................................................................................................................................................................................. 69 7.1
Classification ................................................................................................................................................................................... 69
7.2
Power Requirements................................................................................................................................................................... 69
7.3
Complied Standards .................................................................................................................................................................... 69
7.4
X-ray source assembly............................................................................................................................................................... 70
7.5
Beam Limiting Device ................................................................................................................................................................ 70
7.6
Detector ............................................................................................................................................................................................ 71
7.7
Dimensions and Weights ......................................................................................................................................................... 72
7.8
Movement ........................................................................................................................................................................................ 73
7.9
Other Information........................................................................................................................................................................ 73
7.10 X-ray Tube........................................................................................................................................................................................ 74 7.10.1 OPX – 105 (CEI) ................................................................................................................................................................... 74 7.10.2 D-054SB (Toshiba) ............................................................................................................................................................. 76 8.
Labeling .......................................................................................................................................................................................................... 78 8.1
GDP-1 ................................................................................................................................................................................................. 78
8.2
GDP-1C .............................................................................................................................................................................................. 79
Appendix 1. ........................................................................................................................................................................................................... 82
PAPAYA PLUS User manual
3
Introduction of Manual
Introduction of Manual
GENORAY’s GDP-1C is digital panoramic X-ray system. This manual is designed for your safe use and operation of GDP-1C. When using GDP-1C, use .
The potential risk always exists when using X-ray products and electric medical equipments. All users should be aware of the methods of use, solutions in emergency conditions and other safety articles provided in this manual.
The user must acknowledge the appropriate ways to protect oneself from the potential danger that could occur at any time, and be aware of the condition of the product that can cause danger.
The manufacturer, assembler, and the importer hold responsibility upon the performance of safety, trust and the article below. - Guarantees the installation of the eligible professional, calibration, and repair. GENORAY Co., Ltd is continuously developing the products. The manual or composition of the package might change without notice Tel: +82-31-737-8020
FAX: +82-31-737-801 8
E - mai l: g en ora y@g eno ra y. com Address:
#512
Byucksan
Techn opia
434-6,
Sangdaewon
1-Dong,
Jungwon -Gu,
S eongnam-Cit y, Gyeongg i-Do, Ko rea
4
PAPAYA PLUS User manual
Certificate of Warranty
Certificate of Warranty Range and terms for warranty GENORAY guarantee this product without any expense for the mal-function or disorder occurred under standard transportation and operation within 1 year from installation. Defect on its appearance is out of the range of warranty.
Requirements on After service call When the malfunction or disorder appeared, stop operating and check again the related article of the “User’s Manual.” Before your calling, please put off this equipment and check for the Model Name, serial No. and purchasing date. GENORAY Co., Ltd. is not responsible for indirect harm. GENORAY Co., Ltd. cannot warrant for defect or harm after warranty period.
PAPAYA PLUS User manual
5
Electrical Safety
Symbols used in product
Symbol
description X-Ray Source Assembly, X-ray radiation
Symbol
description
Manufacturer, Address
Caution
Manufacturing Date
Refer to User’s Manual
Product serial number
Type b applied part
WEEE Mark
AC (Exchange)
Laser
This symbol is information about the special matters that require attention while using this software.
This symbol is warning the situation or environments which might bring danger to the patients or users.
6
PAPAYA PLUS User manual
Warning before Use
Warning before Use This machine can be dangerous for all users if the safe exposure condition and appropriate using method are not attended. This machine can only operated by authorized user. This machine need to operate after studying of manual. Random modification on machine is prohibited. There is prohibition of installing different composite parts unless they are supplied by manufacturer. Do not connect equipment’s power source with multi-adaptor or extension cord. This can cause defect or become cause of mal-function. Do not operate machine when the machine is defect or malfunctioning. Please contact engineer or service center for help. Do not use provided PC for unintended purpose. There is probability some problem could happen on transferring images and reconstructing if the PC uses for unintended purpose.
PAPAYA PLUS User manual
7
Electrical Safety
Electrical Safety
The electric circuit of this product uses the electric current which can cause death due to severe injury or electric shock. To avoid this kind of danger, do not remove any of the covering part.
To avoid failure and the danger that might cause serious injury or electric shock to users and patients, please acknowledge the safety procedures listed below. Before cleaning the product, always turn off the power and separate the power cord, and clean the product with lightly wet cotton or sponge. Never let anyone remove the cover except for the professional technician of GENORAY. Do not place food on any part of the product. They might get into the electric circuit and get into contact with electrical matters
The product is not waterproof to water, soap or liquid substances. If the liquid contacts the product, it may cause fire or electric shock. If the liquid was in the product by accident, do not touch the power cord connecting to the power, and wait until the liquid is completely dry.
The user should follow the safety guideline provided below. Turn off the power and separate the power cable from the socket. Stay far away from the place where the product has been installed. Ask for help around you. Use the fire extinguisher authorized only for the electric fire.
Abnormal use of fire extinguisher could be dangerous and cause fire or electric shock. To avoid this kind of danger, check to see if the product can be used where the product has been installed, and always use the authorized product.
8
PAPAYA PLUS User manual
PAPAYA Equipment Introduction Before use this product, please check below.
Before connecting the power, check to see of the power supplying line is appropriate to this product. Check the power and connect the power cable of this product to its power socket. Check the categories mentioned above in case of power input of all accessory (the signal input/output including accessory).
For reducing the risk of electrical shock, and when connecting to power, there needs protecting ground connection.
Place this product away from other X-ray products, power generating equipment and broadcasting stations. If the product is sharing the power with other electric devices, abnormal image might occur.
The protection level of PAPAYA Plus is the 1st grade equipment or type-B equipment. This product does not prevent the invasion of liquid substances.
PAPAYA PLUS User manual
9
PAPAYA Equipment Introduction
Radiation Safety This product has to follow designated country’s radiation law.
This product needs X-ray room. Except the patient, no one can enter X-ray room unless they are patient. While taking exposure, clear space between patient and equipment need to be maintained.
For this product’s safe operation, it needs to be installed at appropriate location with permission of owner. Also, user needs to guarantee about whether equipment will be used at place where owner has chosen.
This product’s owner is, while using this equipment, all users need to be guaranteed about wearing X-ray protection device.
When exposure of X-ray, operator needs to stay away from equipment more than 2 meters.
For this product, users need to wear TLD badge, and check dosage exposure regularly.
When the pregnant woman or young child need to take exposure and need to consult with specialist before.
If you do not follow written rules in this manual, there can be risk for operator and patient.
10
PAPAYA PLUS User manual
PAPAYA Equipment Introduction
1. GDP-1C Equipment Introduction 1.1 Intended Use Digital Panoramic & Cephalometric X-ray System, GDP-1C is used for taking image of anteroposterior’s panoramic view, head’s cephalo view. This equipment is only used for doctor, dentist, and radiologist.
1.2 Equipment’s Characteristics This equipment is used for diagnosing of teeth, chin, and head. This equipment is producing and is controlling X-ray, and composed of X-ray generator, X-ray control part, and X-ray support part. The basic concept of acquiring image is to produce high voltage between X-ray tube’s both positive and negative, and exposure X-ray to subject, and transmitted X-ray to exposure at sensor and acquire the image.
X-ray Generator
High Frequency, Fixed Type Tube, 60 ~ 90kV, 4 ~ 12mA
Focal Spot Size
0.5mm
Image Detector
CdTe flat panel
FOV Image Acquisition / Angle (Standard panoramic)
Panorama: 4.8mm x 150mm Ceph: 4.8mm x 240mm 210°
Scan Time / Exposure Time
Panorama: 17sec(max), Cephalo: 17sec(max)
Patient Position
Standing
Reconstruction Type
Fan beam
Reconstruction Time
Less than 30 seconds
Weight
110 kg ± 5kg (Basic), 140 kg ± 5kg (Cephalo)
Main body
PAPAYA PLUS User manual
1010 x 1055 x 2312 mm (Basic) 1010 x 1775.4 x 2312 mm (Cephalo)
11
PAPAYA Equipment Introduction
1.3 System Basic Specification
GDP-1C
Hand Switch
PC
Ethernet
[ System Connection Image ]
PC is used for acquiring and displaying image, and for image processing.
12
PAPAYA PLUS User manual
PAPAYA Equipment Introduction
1.4 Equipment Composition
This machine has panoramic only model of GDP-1 and has added cephalo exposure that is GDP-1C. Each model’s composition or exterior, please refer to below pictures.
9
3
9
4 6
13 10 5 7 8 11
2
12
1
[ GDP-1 ] No.
[ GDP-1C ] description
No.
description
1
BASE
7
Patient supporting area
2
Fixed column
8
X-ray control equipment
3
Driving Unit
9
Cephalo Arm
4
Rotating Unit
10
On/off switch
5
X-ray source assembly
11
Detector (Panorama, Cephalo using for
12
Emergency stop switch
both, detachable structure)
13
Device status LED
6
PAPAYA PLUS User manual
Hand Switch (Exposure button, Emergency stop switch)
13
PAPAYA Equipment Introduction
1.5 Accessories
Patient’s condition or according to exposure mode, chin rest can be changeable.
[Standard Exposure]
[For No Teeth Patient]
[SINUS & TMJ mode]
Chin rest change method
For appropriate exposure mode, insert correct chin rest to designated hole.
Must use spare part that is provided by manufacturer. If you damage machine when using different spare part, and then the manufacturer does not have responsibility of fault.
14
PAPAYA PLUS User manual
PAPAYA Equipment Introduction
1.6 Patient Apply Part
Cephalo part
Panorama part
Bite
Chin
rest
Temple support
Nasal positional
Patient
handle
Ear rod
After exposure, sanitize patient apply part.
PAPAYA PLUS User manual
15
PAPAYA Equipment Introduction
1.7 On / Off Switch
OFF
ON
Equipment’s on/off switch is like on picture, it is located at equipment’s below part. If turning on equipment’s switch, and the equipment’s LED has light and displays equipment’s waiting condition.
16
PAPAYA PLUS User manual
PAPAYA Equipment Introduction
1.8 Hand Switch Hand switch is made with magnet and can be hung on wall. At hand switch, as seen on picture right, there is X-ray exposure switch and emergency stop switch,
they are
emergency
switch
two functions together. is
used
when
While
equipment
X-ray
exposure
switch
is
malfunctioning or it can harm patient. At moment, equipment will stop.
Emergency stop switch
Exposure switch need to be pushed continuously while is to exposure patient’s diagnostic image. If during exposure, switch is not pushed and machine’s movement is stopped instantly and at OP, advisor word will be displayed.
PAPAYA PLUS User manual
17
PAPAYA Equipment Introduction
1.9 Emergency Stop Switch
Emergency stop switch is located at above picture. When emergency situation happens and push button and equipment’s driving unit or X-ray source assembly’s power supply stops and machine will stop.
When pushing emergency stop switch, turning switch to clockwise to unwind and machine will operate normally.
18
PAPAYA PLUS User manual
PAPAYA Equipment Introduction
1.10
Equipment’s Control Button
Equipment is controlled by equipment’s control button and controls equipment’s movement or controls laser. Each button’s function is as below.
Controlling equipment’s height. ‘LIFT DOWN’ button moves equipment downward and ‘LIFT UP’ button moves equipment upward.
Using when user wants to change the detector.
Turn on/off equipment’s laser. When pushing once, laser turns on and pushing one more time and laser turns off. Laser will turn off after time. It will turn off automatically.
Using when equipment’s gentry is moved forward and backward. By using laser, use to decide patient’s exposure location. ‘FORWARD’ button used for moving gentry to user and ‘BACKWARD’ button used for moving genty, opposite side to user.
PAPAYA PLUS User manual
19
PAPAYA Equipment Introduction
1.11
Laser Control Knob
Laser Control Knob
This equipment, as shown above picture, has laser control know. Before to exposure, when controlling patient’s position, using laser control knob to adjust to Frankfort line.
20
PAPAYA PLUS User manual
PAPAYA Equipment Introduction
1.12 Changing Detector This equipment is according to exposure mode, detector needs to move. When to exposure Panorama Detector’s Location
When to exposure Cephalo Detector’s Location
Detector Detachable Method
A. Pushing detector handle area’s switch.
B. While pushing switch, pull detector downward.
C. By disassembling hook part and pushing detector upward and detector will fall out.
Following above change method and doing reverse way and detector will be attached.
PAPAYA PLUS User manual
21
OP (Operation Panel)
1.13 Device status LED
LED Color
description Blue
Rotating side is at rotation
Yellow
while X-ray exposure
Green
When X-ray exposure, ready
White
Completion of X-ray exposure
Red
22
While taking exposure, error occurs or pushing emergency switch
PAPAYA PLUS User manual
OP (Operation Panel)
2. OP (Operation panel) 2.1 OP Structure
Manual’s used picture can differ from real program’s version.
Patient or uneducated person must not operate OP. Please be aware.
9
1
11
2 3
4
12
5
6
10
7
8
[ Operation Panel Program / Main Screen ]
PAPAYA PLUS User manual
23
OP (Operation Panel)
No. 1 2 3 4 5 6 7 8
Subject Exposure
Time
description Display
Screen Patient’s
Information
Display Screen Exposure
Mode
Choose
Button Exposure Program Choose Button Patient’s
Size
Choose
Quality
Choose
Button Image Button Exposure Condition Display and Change Part Equipment
Movement
Displaying current time. Displaying patient’s information who wait for current exposure. Choosing equipment’s exposure mode. Choosing equipment’s exposure mode by exposure program. Choosing according to patient’s size by pre-set value. Deciding exposure image’s quality. Controlling equipment’s exposure condition manually. Deciding
exposure
condition
and
preparation
Button Part
opinion.
9
Equipment display
Display equipment’s condition
10
Jaw shape selection button
Choosing patient jaw shape and size
11
Order list button
Managing patient information and order list
12
Exposure Program Display
Displaying selected exposure program
completion
and
ease before take an exposure, always look equipment’s display and check whether equipment can take an exposure. About equipment’s condition, please refer to 2.6.
24
PAPAYA PLUS User manual
OP (Operation Panel)
2.2 X-ray Exposure Mode GDP-1 is supporting various types of X-ray exposure mode. At 2.2.1 of Operation Panel Program, at main screen, by pushing wanted x-ray exposure mode and push x-ray button and be able to choose program.
2.2.1Panorama program Exposure Mode
Mode Type Program Standard Panoramic Standard panoramic has standard panorama’s image field of view and Taking exposure.
Panorama mode Orthogonal Panoramic Orthogonal
Panoramic
is
comparison
to
Standard
panoramic
program that takes x-ray exposure with x-ray beam’s angle as 90° of penetration to chin, and produce field of view. At standard panoramic image, teeth is overlapping and effect is decreasing, but exposure area’s opposite side’s shadow of teeth and chin and add an effect.
PAPAYA PLUS User manual
25
OP (Operation Panel)
Exposure Mode
Mode Type Program Bitewing Panoramic Bitewing Panoramic program works at panoramic image and take exposure of bitewing area and take exposure of trajectory of patient. In basic, Bitewing Panoramic is orthogonal panoramic and take exposure with same angle.
TMJ lateral TMJ lateral exposure has function of trajectory and side of temporomandibular
joint.
Taking
exposure
of
both
sides
of
temporomandibular joint and comparing, patient opens mouth and the patient closes mouth to distinguish temporomandibular joint exposure.
TMJ mode
TMJ PA TMJ PA takes exposure of temporomandibular joint’s PA direction and
take
exposure
with
trajectory
direction.
Both
sides
of
temporomandibular joint can take exposure and using as patient opening mouth and closing mouth.
26
PAPAYA PLUS User manual
OP (Operation Panel)
Exposure Mode
Mode Type Program Double Lateral-PA Lateral – PA program takes exposure of temporomandibular joints’ PA
and
Lateral
direction
alternately.
Both
sides
of
temporomandibular joint can take exposure and using as patient opening mouth and closing mouth.
Sinus Midsagittal Sinus midsagittal exposure enables to take exposure of maxillary sinus and use it. The image’s focus is on midsagittal surface.
Sinus mode
Sinus lateral Sinus lateral exposure is like Sinus midsagittal, at maxillary sinus, take a exposure at maxillary sinus’ trajectory. From midsagittal surface’s standard, now focus surface can be toward left or right side.
PAPAYA PLUS User manual
27
OP (Operation Panel)
Exposure Mode
Mode Type Program Sinus PA Sinus PA takes an exposure at PA’s direction of maxillary sinus.
28
PAPAYA PLUS User manual
OP (Operation Panel) 2.2.2Cephalo Program Exposure Mode
Mode Type Program Cephalo lateral Cephalo lateral exposure has at patient’s side, and X-ray beam’s penetration of direction.
Cephalo AP Cephalo AP exposure is that at toward AP’s direction, X-ray penetrate and taking exposure of skull. Cephalo mode
Cephalo PA Cephalo PA exposure is that at toward PA’s direction, X-ray penetrate and taking exposure of skull.
PAPAYA PLUS User manual
29
OP (Operation Panel)
Exposure Mode
Mode Type Program Water’s view Water's view is when the patient is at with Frankfort line and wit h detector surface, and taking exposure with 45 degree angle slo pe and is exposuring area of head.
SMV(Submento vertex) SMV takes exposure of according to patient’s Frankfort line and the surface of detector is leveled with horizon and put patient in this position to take exposure.
Carpus Carpus takes exposure of wrist. In order to do Carpus exposure, using carpus exposure wrist panel and install it and take exposure.
30
PAPAYA PLUS User manual
OP (Operation Panel)
2.3 Choose Patient Size At all exposure program, according to patient’s size and can choose appropriate patient’s size.
Among 4 different size, can choose one size, and according to choice of size of patient, each exposure program has saved X-ray exposure condition with trajectory. At each exposure program, X-ray exposure condition is set up differently.
2.4 Exposure Speed and Change of Image Quality At each exposure program, you can choose different image quality. The image quality can choose between Normal, Fast, HD(High Definition). They are total of 3 types and can choose one.
Normal
Fast
HD ( High Definition)
Fast image has lower image quality and HD has longer image processing time but image.
2.5 Selection of Jaw shape Sharp
Normal Square
Small
Average Large
You can choose different jaw shape and size of patients. There are 9 options by following 3 jaw shape (Sharp, Normal, Square) and 3 jaw size (Small, Average, Large).
PAPAYA PLUS User manual
31
OP (Operation Panel)
2.6 X-ray exposure condition change X-ray condition can be changed by clicking the left or right arrow button of kVp & mA condition value on the x-ray condition display screen like the below image. kVp can be changed 1kVp per 1 click and mA can be changed 0.5mA per 1 click. The exposure time can’t be changed.
The exposure condition value is chosen automatically when choosing patient size & exposure program. If you’d like to change the exposure condition by yourself, you can do like above.
2.7 Order list 2
1
3
4
5
32
PAPAYA PLUS User manual
OP (Operation Panel) Click the order list button, refer to 2.1, then the order list management window will be appeared. Functions of each part are described in below table. No.
Subject
description Displaying patients’ information of registered order by following selected filtration option.
1
Order status filter
Ready: Displaying order information be in ready and wait status. All: Displaying all order information. Done: Displaying order information exposure is finished.
2
Worklist
Loading patients’ information from worklist server.
3
Search
Searching order information.
4
Order information window
Displaying registered order information. Time, status, ID, name and exposure program information are displayed. Managing order. Register: Executing order registration window.
5
Order management button
Select: Selecting registered order of order list for taking image. Modify: Modifying registered order information. Delete: Deleting registered order information. Close: Closing order list and returning to main screen of OP.
This below picture shows a window appears when you click ‘Register’ button, user can input patient information and select exposure program with this window.
PAPAYA PLUS User manual
33
OP (Operation Panel)
2.8 Device condition display screen Device condition display screen shows the current condition of device.
Condition
Description It shows the connection of device & PC. When the device is connected with PC, it turns orange color. It shows the temperature of X-ray generating part of device. If the X-ray generating part is overheated, it turns red color. In this case, please don’t exposures X-ray until the temperature becomes normal. Green: normal, Orange: caution, Red: warning. It shows the connection condition of the device sensor to the panorama or cephalo position. “P” turns violet when the sensor is connected to panorama position. And “C” turns green when the sensor is connected to panorama position.
34
PAPAYA PLUS User manual
Image acquisition
2.9 Option setting
PAPAYA OP program needs to set No.
Name
Explanation
1
General
Display composition or user's convenience need to be set
2
DB Server
Material composition or data management need to be set.
3
DICOM
Outside compatibility or DICOM function need to be set
4
Version
Machine or program version need to be displayed
PAPAYA PLUS User manual
35
Image acqusition 2.9.1 General
Name
Explanation
Language
Setting language
Font
Setting Font
DAP
Setting DAP output measurement
Data Format
Setting data format.
Clinic Name
Save by inputting Clinic Name
Volume control
Controlling machine's sound Setting Multi Focus using or not using
Multi Focus
Number
Setting multi focus' pages (3,5 page setting is abled)
Step
Setting space between images
All save
Saving images that is processed with Multi focus
Automatically setting processing / no processing Auto Process
36
Send DICOM
After taking exposure, automatically send with DICOM format
Test Operation
Setting for exposure test.
R-Mark
Setting exposure images' position mark's printing out / not printing out
Use Cephalo
Adding cephalo mode function (*Option)
Save Raw images
Setting Raw image's setting (If checking and taking up a lot of hard disk
PAPAYA PLUS User manual
Image acquisition
2.9.2 DB Server
Name SQL Server Information
Data folder Information
Explanation Server Name
Setting Order DB's usage path
Triana Server Name
Setting DB path at Triana
Temp Image
At PAPAYA OP, setting saved image path
Triana Image Parameter
Trouble shoot
PAPAYA PLUS User manual
At Triana, by managing image, making im age saving path Setting parameter path
When problem happens, pushing button to solve problem.
37
Image acqusition 2.9.3 DB Server
Name
38
Explanation
Linkage
Setting about interface about compatibility
Equipment Information
Inputting machine's information for DICOM usage
Storage Server
Setting DICOM Storage server
MWL Server
Setting Work list server
DICOM Printer
Setting DICOM Printer
System Log
When pushing test button and printing out processing result
PAPAYA PLUS User manual
Image acquisition 2.9.4 Version
Name
Explanation
PAPAYA OP
Displaying program version at PAPAYA OP
PAPAYA Bridge
Displaying program version of PAPAYA Bridge
PAPAYA REC
Displaying version of PAPAYA REC program
M/B
Displaying Papaya's main board version
sound
Displaying PAPAYA sound controller
LIFT
Displaying PAPAYA LIFT version
Sensor Serial num
Displaying sensor serial number
PAPAYA PLUS User manual
39
Image acqusition
3. Image acquisition 3.1 Registration of patients and orders. 3.1.1 Registration with Triana software.
Please refer the Triana manual provided with device for more details.
Please follow the process below after running Triana software to operate OP for the device operation. Patient registration
Click the patient registration button after running Triana.
Put
patient
patient Click
information
registration
the
patient
in
window.
registration
after done.
A. OP operation [ Panorama ]
Choose the device among device selection tap once the patient registration is done. Papaya is a panoramic device so you can choose panorama. If the case
[ Cephalo ]
of the device including cephalo function, you can choose one between panorama & cephalo. Order information will be registered when you click the acquisition button.
40
PAPAYA PLUS User manual
Image acquisition
3.1.2 Registration with OP software.
Execute order registration input window; Refer to 2.7, Press register button positioned at bottom-left of order list. Input patient information and select exposure program to the window. Order information will be registered when you click the register button.
Then
you
can
see
the
registered
order
information in the order list.
PAPAYA PLUS User manual
41
Image acqusition
3.2 Panorama mode Set position of patient Panorama takes panorama images of upper/lower jaw.
1. Set detector on the exposure position.
Please refer 1.12 about how to install the detector.
2. [ For regular teeth ] Set chin-rest & bite on the device. The patient who doesn’t have teeth or hard to bite [ For regular teeth ]
[For patient without teeth]
the bite should install & use the chin-rest for patient without teeth.
In case of using a chin-rest [For regular teeth], cover it with clean cover. Remove clean cover used and don’t use again.
Refer 1.5 about how to install the chin-rest.
42
PAPAYA PLUS User manual
Image acquisition
Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, necktie for the safety.
3. Operate OP by clicking panorama exposure button after registering a patient according to 3.1 4. When you click “Confirm” button after setting panorama mode & exposure condition on operating OP, “Confirm” button turns “Ready” button, and then patient position arrange window comes out.
You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure change is possible on start screen by ‘Back’ button.
PAPAYA PLUS User manual
43
Image acqusition 5. After being the patient positioned, set height on patient’s height by using 1.9 device set button.
6. Cover bite with clean cover, let the patient put their chin on chin-rest and bite the bite.
If patient can’t raise their neck, set the height of device by using device set button.
7. When the position of patient is correct, lead patient to grab a handle.
8. Set the position of patient by pressing laser button (
) of device set button while the patient bites
the bite correctly.
Laser is consists of 3, you should let the patient be positioned for all 3 lasers correctly to get a good quality of image
9. When the laser is turned on, set midsagittal laser to midsagittal of patient by moving patient’s head.
44
PAPAYA PLUS User manual
Image acquisition
10. Set Frankfort laser to Frankfort of patient by controlling angle of patient’s head. If the position of laser is not correct, set the height of laser by moving Frankfort laser control knob.
Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.
11. Finally move the position of laser by using ‘FORWARD’ & ‘BACKWARD’ button on patient control button to be canine laser positioned on dogtooth of patient.
Laser is turned off automatically after few minutes or if the exposure starts.
12. Check whether the position of 3 lasers is correct after positioning the patient by using Temple support. If it is not correct, be the patient positioned correctly.
PAPAYA PLUS User manual
45
Image acqusition
3.3 TMJ mode patient position set TMJ mode takes the image of lower teeth on TMJ part.
1. Set detector on panorama position.
Refer 1.12 on how to install detector
2. the chin-rest [for TMJ mode]
[ For TMJ mode ]
Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, neck tie for the safety.
46
PAPAYA PLUS User manual
Image acquisition 3. Operate OP by clicking panorama exposure button after registering patient according to 3.1 4. Set TMJ mode & exposure condition on operating OP and click “Confirm” on right-low side, then “Confirm” button turns to “Ready” button, patient position control screen comes out.
You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure change is possible on start screen by ‘Back’ button.
PAPAYA PLUS User manual
47
Image acqusition 5. Set the height by using 1.9 device control button after being the patient positioned.
6. Let the patient lean on the chin-rest like the image left.
Let the patient be positioned according to open/close the mouth.
If patient can’t raise their neck, set the height of device by using device set button.
7. When the position of patient is correct, lead the patient grab the handle.
8. Set the position of patient by pressing laser button (
) of device set button while the patient
stands on correctly. Laser is consists of 3, you should let the patient be positioned for all 3 lasers correctly to get a good quality of image
9. When the laser is turned on, set midsagittal laser to midsagittal of patient by moving patient’s head.
48
PAPAYA PLUS User manual
Image acquisition
10. Set Frankfort laser to Frankfort of patient by controlling angle of patient’s head. If the position of laser is not correct, set the height of laser by moving Frankfort laser control knob.
Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.
In the case of the exposure program is set to PA mode, move the Frankfort laser to the TMJ PA laser marked position and adjust patient’s head angle to position the TMJ is on to the laser line.
11. Finally move the position of laser by using ‘FORWARD’ & ‘BACKWARD’ button on patient control button to be canine laser positioned on TMJ of patient. ‘FORWARD’ button moves to user’s direction but ‘BACKWARD’ button moves to opposite direction of user’s.
Laser is turned off automatically after few minutes or if the exposure starts.
PAPAYA PLUS User manual
49
Image acqusition
12. Check whether the position of 3 lasers is correct after positioning the patient by using Temple support. If it is not correct, be the patient positioned correctly.
50
PAPAYA PLUS User manual
Image acquisition
3.4 Sinus mode patient position set TMJ mode takes the image of lower teeth on TMJ part.
1. Set detector on panorama position.
Refer 1.12 on how to install detector
2. Install the chin-rest for [Sinus mode] on device.
[ for Sinus mode ]
Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, neck tie for the safety.
PAPAYA PLUS User manual
51
Image acqusition
3. Operate OP by clicking panorama exposure button after registering patient according to 3.1 4. Set Sinus mode & exposure condition on operating OP and click “Confirm” on right-low side, then “Confirm” button turns to “Ready” button, patient position control screen comes out.
You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure change is possible on start screen by ‘Back’ button.
52
PAPAYA PLUS User manual
Image acquisition 5. Set the height by using 1.9 device control button after being the patient positioned.
6. Let the patient lean on the chin-rest like the image left.
Let the patient be positioned according to open/close the mouth.
If patient can’t raise their neck, set the height of device by using device set button.
7. When the position of patient is correct, lead the patient grab the handle.
8. Set the position of patient by pressing laser button (
) of device set button while the patient stands
on correctly. Laser is consists of 3, you should let the patient be positioned for all 3 lasers correctly to get a good quality of image
PAPAYA PLUS User manual
53
Image acqusition
9. When the laser is turned on, set midsagittal laser to midsagittal of patient by moving patient’s head.
10. Set Frankfort laser to Frankfort of patient by controlling angle of patient’s head. If the position of laser is not correct, set the height of laser by moving Frankfort laser control knob.
Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.
11. Finally move the position of laser by using ‘FORWARD’ & ‘BACKWARD’ button on patient control button to be canine laser positioned on Sinus of patient. ‘FORWARD’ button moves to user’s direction but ‘BACKWARD’ button moves to opposite direction of user’s.
Laser is turned off automatically after few minutes or if the exposure starts.
54
PAPAYA PLUS User manual
Image acquisition
1. Check whether the position of 3 lasers is correct after positioning the patient by using Temple support. If it is not correct, be the patient positioned correctly.
PAPAYA PLUS User manual
55
Image acqusition
3.5 Cephalo mode patient position set
1. Set detector on cephalo position.
Refer 1.12 on how to install detector
Please let the patient remove eye glasses, hearing aid, false/artificial teeth, hair pin, other accessories and metal etc. before taking X-ray images. If not, the shadow or reflected images can be seen. Please remove the cloths wearing around the neck such as scarf, muffler, and necktie for the safety.
2. Operate OP by clicking cephalo exposure button after registering patient according to 3.1 3. Set mode & exposure condition on operating OP and click “Confirm” on right-low side, then “Confirm” button turns to “Ready” button, patient position control screen comes out.
56
PAPAYA PLUS User manual
Image acquisition
You can’t change the X-ray condition after clicking ‘Confirm’ button. X-ray condition & exposure change is possible on start screen by ‘Back’ button.
4. Set the height by using 1.10 device control button after being the patient positioned. 5. Let the patient positioned according to exposure program after being Ear rod positioned to ear hole of the patient. 6. Set the position of patient by pressing laser button (
) of device set button while the patient bites
the bite correctly. 7. Laser enters horizontally. Control the angle of laser & patient’s Frankfort according to exposure program.
Frankfort plane is the straight line connecting upper part of an earhole & lower part of eyes.
PAPAYA PLUS User manual
57
Image acqusition
3.6 Image taking
After finishing the fixing the position of patient, press ‘Ready’ button of right-low part of OP after informing the patient not to move. After the message for taking a X-ray image from OP, the exposure starts when pressing X-ray shot switch.
Users should exposure from the position safe from X-ray.
Please keep pressing the switch during exposure. If the users take their hands off from shot switch, the exposure is stopped and then, the error message comes out about the abnormal shut down. Focus on patient & condition of device during exposure. If the emergency situation is occurred, take the hands off right away and then press emergency switch
58
PAPAYA PLUS User manual
Image acquisition When the exposure starts, OP shows X-ray warning message & the image taking, and final image comes out again after the complete.
PAPAYA PLUS User manual
59
Image acqusition
3.7 Image Control
After image exposure finishes and on screen, image is displayed and before push "save" button and can do image control
Drag mouse's left button
Move image within window (Up/down/left/right)
Drag mouse's right button
Control Brightness
Drag mouse's both left and right
View enlarge or minimize
Double click mouse's left button
Re-size image with size of window
Double click mouse's right button
initialize image to before control brightness
Move mouse wheel to up or down
60
View if images are many and previous or next i mage
PAPAYA PLUS User manual
Image acquisition
3.8 TWAIN INTERFACE If user's program supports Twain interface, and by using TWAIN data source in which is named, "PAPAYA" (TWAIN Version 2.1) and can send image to user's program. Sending methods are supported with Native Mode Transfer and Buffered Memoray Mode Transfer. The image acquisition by using PAPAYA data source is following. 1. Connect TWAIN session by choosing "PAPAYA" among TWIN data source whithin user's program 2. At user's program, the command of Acquire and send this command, and progress on OP's full screen mode. 3. From contents of 3.2 to 3.6, depending upon exposure mode and choose exposure mode and control patient position, and then try exposure of X-ray. 4. After exposure steps are finished normally and click "save" button and send user's program called, "PAPAYA." 5. After sending is completed, and OP screen is closed automatically and ends with TWAIN session. TWAIN data source can send image according to usage of Multi Focus that enables to send more than 1 page. If on user's program and try to receive 1 page of image, and un-check the check mark on content of 2.9.1's at general option tab and un-check All Save's checkbox.
3.9 OP Additional Function OP enables burning CD, sending to DICOM server, registering additional command, opening of patient date folder. Burning CD and sending of DICOM server function is able for patient who completed taking exposure. On order list window, clicking mouse's right button and by choosing patient and pop-up menu appears for choosing additional function.
PAPAYA PLUS User manual
61
Image acqusition 3.9.1 Sending to DICOM Server At pop-up menu, by choosing send, and this function allows to pre-set to send image to DICOM server. The DICOM server information is written on 2.9.3 and on DICOM option tab's lower left side and set on Storage Server. By pushing "Insert Item" button, and input information window appears and input AETitle, IP, Port, and Timeout and push "Insert" button. At Storage Server's Preset list, and choose server in which will be sent and finish job by pushing "Save" button.
3.9.2 Burning CD After taking exposured image and save them by using CD-ROM like media storage device or hard disk. On pop-up menu, choose "Save to CD" and progress CD burning window.
For media storage device, choosing right side at device list and save image. Checking by Total Space Used and Free Space and click "Burning" button. After burning is completed and finish and click "Close" button. When saving at hard disk or USB, by pushing
button and choose save path and process burning.
With image, image viewer of CDSee is also saved. Media device is connected, but device list doesn't appear and can't choose device, and push "Refresh" button for updating the device list On same path inside of CDSee folder, progress CDSee.exe and view saved image.
62
PAPAYA PLUS User manual
Image acquisition
3.9.3 Register addition order Additional order function is that on previous ordered list, and patient need additonal exposure and doesn't input basic personal information. Then choose exposure program and do registration for order. If you choose additional order, and have chosen patient's basic information is updated and progress with registration ordered window. By choosing exposure program, and click "Register" button, and the order is added on ordered list.
3.9.4 Opening of patient's folder When choosing opening of folder, search window is opened by starting from where the images are in saved folder.
PAPAYA PLUS User manual
63
Image acqusition
Use condition Please avoid these conditions for the operation/storing of this device.
Exposure to humidity Exposure to Exposure to dust Exposure to high humidity Exposure to condition where has problem with ventilation Exposure to salty air. Exposure to chemical / harmful gas.
Please avoid use the product the place where has strong vibration or over heat or humidity.
Condition of movement & Storing Temperature : -20°C ~ 60°C Relative humidity: 5% ~ 95% RH (But, No dew condensation) Air pressure : 500 ~ 1100 hPa
Condition of movement when using Temperature: 15°C ~ 30°C
Relative humidity : 5% ~ 75% RH (But, No dew condensation.)
64
PAPAYA PLUS User manual
Message
4. Cleaning You can clean Bite piece, patient support table, chin rest & patient handle by high pressure sterilization up to 135℃ or cotton/gauze with alcohol.
You can patient handle with cotton/gauze with alcohol.
Please clean softly OP panel or other parts with wet soft cloth.
Clean the device after turning off the power.
Don’t spray liquid type of cleaning product directly to the device.
5. Handling of waste It is regular custom to follow the law of each country when handling of waste including devices. Please follow the regulations of each country.
PAPAYA PLUS User manual
65
Message
6. Message 6.1 Error message Please refer the solutions below when the messages like below come out on OP. If, below methods do not solve problems and please contact manufacturer
1
Error message Temporary Solution Error message
2
Temporary Solution Error message
3
Temporary Solution Error message
4
Temporary Solution Error message
5
Temporary Solution Error message
6
Temporary Solution Error message
7
Temporary Solution Error message
8
Temporary Solution Error message
9
Temporary Solution Error message
10
Temporary Solution Error message
11
66
Temporary Solution
kV Low Occurred Please turn off machine and turn on. kV High Occurred Please turn off machine and turn on. mA Low Occurred Please turn off machine and turn on. mA High Occurred Please turn off machine and turn on. OCP Please turn off machine and turn on. EPIP Please turn off machine and turn on. Tank Overheat (SW) Turning off machine for one hour and use again Tank Overheat (sensor) Turning off machine for one hour and use again Exposure button was released If user press off button by mistake, and after 10 seconds, error message disappear and processing again Interrupt by something to rotate gantry If user touched rotating part while rotation, after 10 seconds, and error message disappear and use again Please, recheck laser alignment If Canine Laser is located at back of ear, and after 10 seconds and error message disappeared and use again.
PAPAYA PLUS User manual
Message
Please contact to manufacturer when the error messages like below.
No.
Error message
1
Gantry rotating overload
2
Y-axis moving overload
3
Ceph detector moving overload
4
Ceph collimator moving overload
5
Collimator moving overload
6
No program at FPGA
7
Exposure program was wrong.
PAPAYA PLUS User manual
67
Message
6.2 Condition message No.
68
Message
설 명
1
Scanning..
Taking a image during rotating
2
Processing
3
Testing
Testing the rotating part.
4
Make by Ready
Heating a X-ray tube.
5
Press the switch
Exposure X-ray pressing the switch.
6
Returning...
Returning the rotating part to starting point.
7
Emergency
Pressing emergency button.
8
Technical Mode
Setting OP system from PC.
Making a quality of image better. Reorganizing the image for diagnosis.
PAPAYA PLUS User manual
Technical Data
7. Technical Data 7.1 Classification Electrical classification
Class I, Type B
7.2 Power Requirements Input Power Supply
220 V~
Power Supply Frequency
50/60 Hz
Capacity of Power Supply
2.2 kVA
You are recommended to use AC power to avoid crash of power requirements for other devices.
7.3 Complied Standards
69
- IEC/EN 60601-1
- IEC/EN 60601-1-2
- IEC/EN 60601-1-3
- IEC/EN 60601-2-7
- IEC/EN 60601-2-28
- IEC/EN 60601-2-32
User manual
Technical Data
7.4 X-ray source assembly kV range
60~90 kV
mA range
4~12 mA
Exposure time range
PANORAMA: max 17sec CEPH: max 17sec Inherent Filtration : 0.5 mmAl (OPX-105) Inherent Filtration : 0.8 mmAl (D-054SB)
Total filtration
Added Filtration : 2mmAl Total filtration : 2.5mmAl (OPX-105) 2.8mmAl (D-054SB)
Superheated Temperature
50 ℃
7.5 Beam Limiting Device
70
Material
Aluminum + 3.0 mm Pb
Form
Rectangle
PAPAYA PLUS User manual
Technical Data
7.6 Detector XID-C15DP (Only for Panorama) Items
Specifications
Sensor Material
CdTe, 750 um
Pixel Size (um)
100
Active Area (mm)
150 mm x 4.8 mm
Dynamic Range (dB)
≥ 72dB
DQE(70kV,0lp,mm)
≥ 0.8
MTF(2lp,mm)
≥ 0.8
Power Supply Requirement
Digital +5 V/1 A, Analog +5 V/1 A
Command I/O
3 TTL trigger
Data I/O
GigE
Frame Rate (fps)
200 fps
Data Bits Per Pixel
12 bit
Mechanical
Size (mm)
73W*168H*21.05D
Dimension
Weight (g)
340
Part
Sensor Chip
Electronics
XID-C24DC (For Panorama and Cephalo) Items
Specifications
Sensor Material
CdTe, 750 um
Pixel Size (um)
100
Active Area (mm)
240 mm x 4.8 mm
Dynamic Range (dB)
≥ 72dB
DQE(70kV,0lp,mm)
≥ 0.8
MTF(2lp,mm)
≥ 0.8
Power Supply Requirement
Digital +5 V/1 A, Analog +5 V/1 A
Command I/O
3 TTL trigger
Data I/O
GigE
Frame Rate (fps)
Pano: 200 fps, Ceph: 125 fps
Data Bits Per Pixel
12 bit
Mechanical
Size (mm)
73W*255H*21.7D
Dimension
Weight (g)
505 g
Part
Sensor Chip
Electronics
PAPAYA PLUS User manual
71
Technical Data
7.7 Dimensions and Weights GDP-1 치
수
1040 x 1155 x 2317 mm
중
량
110 kg ± 5%
GDP-1C
72
치
수
1040 x 1775 x 2317 mm
중
량
140 kg ± 5%
PAPAYA PLUS User manual
Technical Data
7.8 Movement
Up/Down Movement
Right/left Movement
Rotating Angle
Movement (Distance) : 700 mm Speed : 16.7 mm/s Movement (Distance) : 170 mm Speed : 6.8 mm/s 220 °
7.9 Other Information Maximum No. of Exposure (Per an hour)
10 times
X-ray Exposure method
Continuous
SID
SSD Maximum tube current at maximum tube voltage Maximum tube voltage at maximum tube current Tube current and voltage at maximum power output
Panorama : 480 mm Cephalo: 1650 mm Panorama : 248 mm Cephalo : 1359 mm 12 mA 90 kV 90 kV, 12 mA
Minimum mAs
36 mAs
Over current Release
Fuse 15A
Ground connection
B type Ground
PAPAYA PLUS User manual
73
Technical Data
7.10 X-ray Tube 7.10.1 OPX – 105 (CEI) Type
Stationary
Focal Spot
0.5 mm
Target Angle
5°
Target Material
Tungsten
Inherent Filtration
0.5 mmAl
Cooling Method
Oil cooling
Nominal Anode Input Energy at 0.1S(DC)
2000 W
Nominal Voltage
110 kV
Thermal Characteristics Anode Heat Storage Capacity
30 kJ
Maximum Anode Cooling rate
200 W
X-Ray tube characteristic curve (1) Maximum Rating Charts (Absolute maximum rating charts)
74
PAPAYA PLUS User manual
Technical Data (2) Emission Characteristics
(3) Anode Thermal Characteristics
(4) OPX/105 size (Unit:mm)
PAPAYA PLUS User manual
75
Technical Data 7.10.2 D-054SB (Toshiba) Type
Stationary
Focal Spot
0.5 mm
Target Angle
5°
Target Material
Tungsten
Inherent Filtration
0.8 mmAl
Cooling Method
Oil cooling
Nominal Anode Input Energy at 0.1S(DC)
1750 W
Nominal Voltage
110 kV
Thermal Characteristics Anode Heat Storage Capacity
35 kJ
Maximum Anode Cooling rate
250 W
X-Ray tube characteristic curve (1) Maximum Rating Charts (Absolute maximum rating charts)
76
PAPAYA PLUS User manual
Technical Data
(2) Emission Characteristics
(3) Anode Thermal Characteristics
(4) D-054SB size (Unit: mm)
PAPAYA PLUS User manual
77
Labeling
8. Labeling 8.1 GDP-1
• Digital Sensor
• X-ray tube • Beam limiting device • Generator • Rotating Arm • Hand S/W
• Laser
• Manufacturer Label • X-ray Support
78
PAPAYA PLUS User manual
Labeling
8.2 GDP-1C
• Digital Sensor
• X-ray tube • Beam limiting device • Detector
• Generator • Rotating Arm • Laser
• Hand S/W
• Manufacturer Label • X-ray Support
PAPAYA PLUS User manual
79
Labeling [Manufacturer label]
- GDP-1
- GDP-1C
80
PAPAYA PLUS User manual
Labeling
PAPAYA PLUS User manual
81
Maintenance Appendix 1.
MAINTENANCE
The user needs to check that all safe related the equipment operation before using it. Routine maintenance - Machine and parts should be maintained by basic condition. - Machine and parts should be checked by routine maintenance. Regular maintenance - Check list every 1 year Accuracy of X-ray tube voltage Accuracy of X-ray tube current Exposure dose of patient Maintenance for long break If the machine is not used for 6 months, the machine should be inspected by manufacturer. - Check list Accuracy of X-ray tube voltage Accuracy of X-ray tube current Exposure dose of patient
It must be used by a trained professional. This machine is maintained by GENORAY America.
Test method - Accuracy of X-ray tube voltage Test instrument: X-ray multi meter Test method: Measure the tube voltage value and compare setting voltage with measured voltage. Test limit: The error of the values of the X-ray tube voltage shall be not greater than 10% - Accuracy of X-ray tube current Test instrument: Oscilloscope Test method: Measure the tube current value and compare setting current with measured current. Test limit: The error of the values of the X-ray tube current shall be not greater than 20%
82
PAPAYA PLUS User manual
Maintenance - Exposure dose of patient Test instrument: X-Ray multi-meter Test method: Measured dosimetry. Test limit: less than 7 mGy (Panorama), 3 mGy (skull AP/PA), 1.5 mGy (skull lateral)
If maintenance parameter is over the limit, some part is changed or repaired in manufacturer’s judgment.
INSPECTION
INSPECTION ITEM
PERIOD
INSPECTOR
RESULT 1
2
3
4
5
Remarks
On / Off Switch Up/Down, Right/Left buttons EVERY DAY
Exposure Hand Switch Patient Registration
User
Image Save and Load Temperature & Humidity EVERY WEEK Emergency Button Tube Voltage Accuracy Test EVERY 1YEARS
Tube Current Accuracy Test Exposure Dose to the patient
WHEN THE MACHINE HAS NOT BEING USED FOR 6 MONTH
Manufacturer Tube Voltage Accuracy Test Tube Current Accuracy Test Exposure Dose to the patient
PAPAYA PLUS User manual
83
GENORAY Co., Ltd. #512 Byucksan Technopia 434-6 Sangdaewon 1-Dong Jungwon-Gu Seongnam-City GyeonggiDo Korea, 462-716 TEL : +82-31-737-8020
FAX : +82-31-737-8018
www.genoray.com
GENORAY America Inc. 1073 N. Batavia St. Orange, CA 92867 Tel: +1-855-436-6729 Fax : +1-714-786-8919
Obelis s.a. Boulevard Général Wahis 53, 1030 Brussels, BELGIUM Tel: +32-2-7325954
Fax: +32-2-7326003
The original version of PAPAYA is being written in Korean
84
PAPAYA PLUS User manual